Research on non-invasive monitoring system based on cerebral blood oxygen on near infrared spectroscopy
DOI:
Author:
Affiliation:

Clc Number:

TN21;R443 + . 8

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    At present, most of the non-invasive monitoring equipment for cerebral blood oxygen in clinical use is dual-wavelength. Because the absorption of melanin will cause deviations in the measurement results, the measurement results need to be corrected. For this reason, based on the basic principle of non-invasive monitoring of cerebral blood oxygen by near infrared spectroscopy, in view of the shortcomings of existing non-invasive monitoring equipment for cerebral blood oxygen, considering the optical characteristics of the prefrontal lobe, a four-wavelength detection light source (700, 760, 805, 850 nm) and a dual-channel photoelectric detector for noninvasive monitoring of cerebral blood oxygen, and a mathematical model of spectral absorbance is established. On this basis, a new type of non-invasive monitoring system for cerebral blood oxygen is constructed, which can inhibit the interference of skin melanin components to non-invasively measure the cerebral blood oxygen information. Finally, through the rank sum analysis of the Valsalva exercise and noload control experiments, the effectiveness of the system is preliminarily verified.

    Reference
    Related
    Cited by
Get Citation
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online: March 06,2023
  • Published: