Abstract:Invasive electrode-based measurement methods are widely used in conductivity measurement systems. However, due to issues such as electrode corrosion and polarization effects, scholars at home and abroad have conducted extensive research on non-invasive electromagnetic conductivity measurement methods. Most of these electromagnetic conductivity measurement methods rely on traditional fixed-frequency amplitude-modulated measurement circuits, which have complex signal processing circuits and are susceptible to noise interference, resulting in measurement errors. Some scholars also measure electrical conductivity by frequency, yet they fail to adopt proper measurement methods and excitation frequencies, resulting in poor measurement performance and low detection system sensitivity. To address the above-mentioned problems, a high-sensitivity non-contact differential frequency solution conductivity detection system is designed. Theoretically, it is analyzed that the eddy current in the electrolyte solution will affect the impedance of the coil. Increasing the excitation frequency and immersing the detection coil into the solution can significantly improve system sensitivity. The detection coil is waterproofed to avoid direct contact with the solution. The detection system converts the change in coil inductance into a change in frequency, and measures the conductivity of the solution by detecting the differential frequency of the coil. This differential structure can reduce the impact of external interference on the detection system. The feasibility of this measurement method was verified through calibration experiments. The detection system achieves a high sensitivity of 54 485 Hz/(mS·cm-1) when measuring low-conductivity solutions, with an overall average sensitivity of 34 340 Hz/(mS·cm-1). Based on the calibration experiment, the fitting equation between conductivity and differential frequency was derived. By comparing the results of the accuracy experiment and the calibration compensation experiment, it can be seen that the relative error of most measurements is within 5% after calibration compensation. Compared with that before calibration, the overall error after calibration is reduced, which reflects the importance of calibration compensation before measurement. The experimental objects are two electrolyte solutions, namely NaCl and KCl. The results indicate that the electrical conductivity of both solutions exerts an influence on the differential frequency, which proves the reliability and universality of the conclusion.