智能总辐射测量仪动态校准方法研究
DOI:
CSTR:
作者:
作者单位:

1.中国气象局智能气象观测技术重点开放实验室北京100081;2.新疆维吾尔自治区气象技术装备保障中心 乌鲁木齐830002;3.中国气象局气象探测中心北京100081;4.临近空间环境特性及效应全国重点实验室 北京100081;5.中国气象局气象探测工程技术研究中心北京100081;6.中国民航工程咨询有限公司 北京100621;7.华云升达(北京)气象科技有限责任公司北京102299

作者简介:

通讯作者:

中图分类号:

TH765;P414.5+1

基金项目:

中国气象局智能气象观测技术重点开放实验室开放研究课题项目(ZNGC2024MS06)、国家重点研发计划(2023YFF0312003)项目资助


Research on dynamic calibration method of intelligent pyranometer
Author:
Affiliation:

1.Key Laboratory of Intelligent Meteorological Observation Technology, China Meteorological Administration, Beijing 100081, China; 2.Meteorological Technology and Equipment Support Center of Xinjiang Uygur Autonomous Region, Urumqi 830002, China; 3.Meteorological Observation Center of China Meteorological Administration, Beijing 100081, China; 4.State Key Laboratory of Environment Characteristics and Effects for Near-space, Beijing 100081, China; 5.Research Centre on Meteorological Observation Engineering Technology of China Meteorological Administration, Beijing 100081, China; 6.China Civil Aviation Engineering Consulting Co., Ltd., Beijing 100621, China; 7.Huayun Sounding (Beijing) Meteorological Technology Co., Ltd., Beijing 102299, China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    智能总辐射测量仪的校准是确保其测量准确性的关键环节。传统校准方法采用校准时段内获得的单一固定修正系数K对仪器测量数据进行修正,而未能考虑仪器响应度随太阳天顶角θZ变化的非线性特性,导致在全天尤其上午和下午天顶角较大时段存在显著的系统性测量偏差。利用DFN5型智能总辐射测量仪,通过设计比对校准实验,采用多项式和分段线性等统计分析方法,构建了 K-θZ动态修正系数模型。通过比较K-θZ动态修正系数和固定系数对独立数据集的验证分析,结果表明,与传统固定系数修正方法相比,基于K-θZ模型的动态修正方法在上午非校准时段的测量中,将太阳总辐射测量数据的平均偏差误差(mean bias error, MBE)由-5.57 W·m-2改善至-2.41 W·m-2(改善约56.7%),均方根误差(root mean square error, RMSE)从6.84 W·m-2减小到3.07 W·m-2(改善约55.1%),t-统计量由28.77下降至26.01;在下午非校准时段,将太阳总辐射测量数据的MBE从-0.60 W·m-2改善至-0.03 W·m-2(改善约95.0%),将RMSE从7.84 W·m-2降低至1.24 W·m-2(改善约84.2%),t-统计量从1.64降低至0.57,有效提高了太阳总辐射测量数据的准确性和一致性。

    Abstract:

    Calibration of intelligent pyranometer is a crucial step to ensure the accuracy of their measurements. Traditional calibration methods use a single fixed correction factor obtained during the calibration period to correct the pyranometer’s measurement data, without considering the nonlinear characteristics of the instrument’s response to the solar zenith angle. This leads to significant systematic measurement errors throughout the day, especially during the morning and afternoon when the solar zenith angle is large. In this study, a DFN5 intelligent pyranometer was used, and a comparison calibration experiment was designed. By applying statistical analysis methods such as polynomials and piecewise linear functions, a K-θZ dynamic correction factor model was constructed, which relates the correction factor K to the solar zenith angle θZ. Through the verification analysis of independent data sets using the K-θZ dynamic correction factor and the fixed factor, the results show that compared with the traditional fixed factor correction method, the dynamic correction method based on the K-θZ model improved the MBE(mean bias error) of the global solar radiation measurement data from -5.57 W·m-2 to -2.41 W·m-2(an improvement of approximately 56.7%) in the morning non-calibration period, reduced the RMSE(root mean square error) from 6.84 W·m-2 to 3.07 W·m-2(an improvement of approximately 55.1%), and decreased the t-statistic from 28.77 to 26.01. In the afternoon non-calibration period, it improved the MBE from -0.60 W·m-2 to -0.03 W·m-2(an improvement of approximately 95.0%), reduced the RMSE from 7.84 W·m-2 to 1.24 W·m-2(an improvement of approximately 84.2%), and decreased the t-statistic from 1.64 to 0.57, effectively enhancing the accuracy and consistency of the global solar radiation measurement data.

    参考文献
    相似文献
    引证文献
引用本文

殷星晨,崇伟,蔡震坤,郭凤娟,马静,吴宁,江丹丹,李楠.智能总辐射测量仪动态校准方法研究[J].电子测量与仪器学报,2026,40(9):70-78