空气负离子测量仪校准方法研究
DOI:
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

TN06

基金项目:

华东区域气象科技协同创新基金合作项目(QYHZ201806)资助


Research on calibration method of air negative ion measurement
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    为解决空气负离子测量仪器校准问题,在简要分析空气负离子测量仪工作原理的基础上,提出了仪器量值溯源途径和 校准设备的研制方法,并对研制的负离子发生装置和标准负离子测量仪进行了试验验证。 验证结果表明,负离子发生装置产生 的负离子场具有负离子浓度连续可调、稳定性好、响应速度快、离子粒径小的特点。 标准负离子测量仪相对误差不超过±5%,标 准装置测量不确定度为 6%,可用于空气负离子测量仪的校准和测试。 利用负离子发生装置和标准离子测量仪组成的校准装 置,对两种不同的负离子测量仪进行了校准试验,对测得的测量误差进行了分析,验证了校准方法的科学合理性和可行性。

    Abstract:

    In order to solve the calibration problem of air negative ion measuring instruments, based on the working principle of air negative ion monitor, a method for tracing the source values of instruments and a method of developing the calibration equipment are proposed. The developed negative ion generator and standard negative anion measuring instrument are tested and verified. The verification results show that the anion field produced by the anion generator has the characteristics of continuously adjustable anion concentration, good stability, fast response speed and small particle size, the relative error of the standard negative ion measuring instrument is not more than ± 5%, and the measurement uncertainty of the standard device is 6%, which can be used for the calibration and test of the negative ion measuring instrument. Using a calibration device composed of the negative ion generating device and the standard ion measuring instrument, a calibration test is performed on two different negative ion measuring instruments, and the measurement errors are analyzed to verify the scientific rationality and feasibility of the calibration method.

    参考文献
    相似文献
    引证文献
引用本文

罗 昶,杨安良,丁妙增,黄鹏良,葛永华,章 焕.空气负离子测量仪校准方法研究[J].电子测量与仪器学报,2020,34(7):89-94

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2023-11-20
  • 出版日期:
文章二维码